Philips HeartStart OnSite AED Defibrillator - Product Specifications

Find detailed specifications about the capabilities of the Philips HeartStart OnSite AED Defibrillator.

 

Defibrillator

Defibrillator Model

HeartStart OnSite (M5066A)

How Supplied

Defibrillator, Owner's Manual, Battery, 1 Adult SMART Pads Cartridge, Quick Reference Guide, Quick Start poster, Carry Case.

Waveform

Truncated Exponential Biphasic. Waveform parameters adjusted as a function of patient impedance.

Energy

Single energy output.
Adults: nominal 150 Joules into a 50 ohm load.
Infant/Child: nominal 50 Joules into a 50 ohm load.
Automatically set based on type of SMART Pads cartridge installed.

Shock-to-Shock Cycle Time

Typically less than 20 seconds between shocks in a series.

Quick Shock

Able to deliver a shock after the CPR pause typically in eight seconds.

Voice Instructions

Detailed voice messages guide responder through use of the defibrillator.

CPR Voice Coaching

Voice instructions in adult and infant/child CPR available at user's option.

Shock Delivery

Via adhesive pads placed on patient's bare skin as illustrated on pads.

Controls

Green SMART Pads cartridge handle, Green On/Off button, Blue i-button, Orange shock button.

Indicators

Ready light; Blue i-button; Caution light.

Physical Specifications

Size
    Height
    Width
    Depth


2.8 inches (7 cm)
8.3 inches (21 cm)
7.4 inches (19cm)

Weight

3.3 lbs (1.5 kg) With battery and pads cartridge

2.4 lbs (1 kg) Without battery or pads cartridge

Environmental/Physical Requirements

Sealing

Drip proof per EN60529 class IPX1.
Solid Objects per EN60529 class IP2X.

Temperature

Operating: 32º - 122º F (0º - 50º C)
Standby: 50º - 109º F (10º - 43º C)

Humidity 

Operating: 0% to 95% relative, non-condensing
Standby: 0% to 75% relative, non-condensing

Altitude

Operating: 0 to 15,000 feet
Standby: 0 to 8,500 feet >48 hours
and 8,500 to 15,000 feet <48 hours

Shock/Drop Abuse

Withstands 1 meter drop to any edge, corner or surface.

Vibration

Meets EN1789 random and swept sine, road ambulance specification in operating and standby states.

EMI (Radiated/Immunity)

Meets EN55011 Group 1 Level B Class B and EN61000-4-3.

Patient Analysis System

Patient Analysis

Evaluates patient ECG and signal quality to determine if a rhythm is shockable and evaluates connection impedance for proper defibrillation pad contact. Rhythms considered shockable are ventricular fibrillation (VF), and certain ventricular tachycardias (VT) associated with lack of circulation, including ventricular flutter and polymorphic ventricular tachycardia. For safety reasons, some VT rhythms associated with circulation will not be interpreted as shockable, and some very low-amplitude or low-frequency rhythms will not be interpreted as shockable VF.

Sensitivity/Specificity

Meets AAMI DF80 guidelines and AHA recommendations for adult defibrillation(Circulation 1997;95:1677-1682).

Artifact Detection

The effects of pacemaker artifact and electrical noise are minimized with artifact detection.

Battery (M5070A)

Type

9 Volt DC, 4.2 Ah, composed of disposable long-life lithium manganese dioxide primary cells.

Capacity

Minimum 200 shocks or 4 hours of operating time.

Install-By Date

Battery is labeled with an install-by date of at least five years from date of manufacture.

Standby Life

Four years typical when battery is installed by the install-by date.(Will power the AED in standby state within the specified standby temperature range, assuming one battery insertion test and no defibrillation uses.)

SMART Pads

Adult SMART PadsCartridge

M5071A defibrillation pads for patients 8 years of age and older or 55 lbs. (25 kg) and over.

Infant/Child SMART Pads Cartridge

M5072A defibrillation pads for patients under 8 years of age or 55 lbs. (25 kg).

Energy Delivered

Adult: nominal 150 Joules into a 50 ohm load.
Infant/Child: nominal 50 Joules into a 50 ohm load.

How Supplied

 

Disposable cartridge, containing adhesive defibrillation pads, clicks into defibrillator for an integrated pads solution.

Active Surface Area

13.2 in2 (85 cm2 ) each

Cable Length

Adult pads: 54 inch (137.1 cm)
Infant/Child pads: 40 inch (101.6 cm)

Use-By Date

Cartridge is labeled with a use-by date of at least two years from date of manufacture.

Training Pads

Adult Training Pads Cartridge

M5073A

Infant/Child Training Pads Cartridge

M5074A

Function

Special pads place HeartStart OnSite into training mode and disable its energy delivery capability. Feature 8 real-world training scripts.

Automated and User Activated Self Tests

Daily Automatic Self-tests

Tests internal circuitry, waveform delivery system, pads cartridge and battery capacity.

Pads Integrity Test

Specifically tests rediness- for use of pads (gel moisture).

Battery Insertion Test

Upon battery insertion, extensive automatic self-tests and user-interactive test check device readiness.

Status Indicator

Blinking green Ready light indicates ready for use.

Data Recording and Transmission

Infrared

Wireless transmission of event data to personal computer using the IrDA protocol.

HeartStart Event Review Software

Data management software (optional) for download and review of data retrieved through defibillator's infrared data port.

Data Stored

First 15 minutes of ECG and the entire incident's events and analysis decisions.